Services - 5.1 Compressed air

All lab2lab and building2building devices except the Alarm require compressed air. 

Compressed air must be dry and oil free (Instrument grade), meeting ISO 8573-1:2010 or better. 

  • water content : Class 4 or better 
  • particulates: Class 2 or better 
  • oil: Class 2 or better 

Each compressed air outlet must have a continuous flow rate of 300SLPM (100NLPM) @ 6bar. 

Only one peripheral device is ever active at a time, unless the system is integrated with a building2building system. 

If compressed air is not available, N2 may be used instead. Please contact SPT Labtech before choosing this option to discuss safety concerns. 

Requirement 

Quantity 

Supply pressure 

6-10 bar allowable, 6 bar preferred 

Flow required when sending a vial 

300 L/min free air (Senders, Receivers, Buffer, Router, Transceiver) 

Flow required when receiving a vial 

300 L/min free air (Router, Transceiver) 

Operating time 

2-10 seconds per device, up to 180 seconds for the Transceiver 

Fitting size 

6mm OD quick release to be provided within 5m of device. 
Device will be supplied with 6mm OD / 4mm ID pipe 

Each device is supplied with a quick disconnect, auto-closing connector. However, SPT Labtech recommends that a local isolator (1/4 turn valve) is provided locally to the devices to enable 
safe isolation of supply air if required. 

5.1.1 Using nitrogen

The use of nitrogen to transport vials is technically acceptable and the requirements are 
identical to those for using compressed air. Normal use of lab2lab should not increase the concentration of nitrogen to unsafe levels assuming suitable ventilation is provided in the environment where it is installed. 

However, nitrogen is poisonous in high concentrations and lab2lab is not designed to detect leakage of compressed gas. While the lab2lab system should be safe, it is essential that if 
nitrogen is used suitable alarm and protection equipment is also installed to warn of high concentrations of nitrogen in the event of a failure. Typically, this is achieved by using an O2 monitor which trips if the O2 concentration drops below 20%. 

5.1.2 Noise levels due to compressed air

The following table gives typical noise levels produced by the equipment in operation. The duration of the noise is dependent upon distances between devices and may vary for each installation. 

Device 

SPL (dBA) 

Duration 

Router 

80 

Between Router and Receiver, Collection station and Buffer typically 3 seconds or less 
Between Router and Sender between 3 and 12 seconds 
Between Router and building2building 1 second 

Buffer 

60 

Typically < 2 seconds 

Bin 

<40 

Only operates a solenoid valve, no airflow 

Sender 

<50 

Up to 12 seconds 

Receiver (all types) 

<50 

Typically < 3 seconds 

building2building Transceiver 

<75 

Up to 2 minutes at receiving end and 10 seconds at sending end 

Transport pipes 

<50 

Almost imperceptible and very brief and occasional 

 

Services - 5.2 Power >>>