Errors and warnings

WARNINGS

Warnings appear when the designer detects a potential issue that may occur when using the
dragonfly. It is still possible to create the CSV file even though these warnings are shown.

warning  Warning – Required diluent volume is [value] μl, which exceeds syringe capacity (4000μl).

Use syringe splitting, adjust stock concentrations, reduce well volume, or raise the gradient
concentration of one or more components.

warning Warning – Required volume for [component name] ([value] μl), exceeds capacity (4000μl).

Use syringe splitting, adjust stock concentrations, reduce well volume, or lower gradient
concentration.

ERRORS

error Gradient concentration too high for component [component name], increase the stock
concentration or reduce the gradient concentration.

error Overflow – total volume of components in at least one well exceeds well volume. Please
adjust gradients or concentrations.

error One or more dispense volumes for component [component name] are less than the
minimum ([value] μl). Increase well volume or reduce min volume.