Packing list
- VolumeCheck instrument
- Flash drive containing the VolumeCheck software and the VolumeCheck manual
- USB cable (1)
- Power Cable (1)
- Rack adapters (per order)
VC100 & VC384 components
Instrument set up
- Remove the Shipping Screws.
- The VolumeCheck models have two shipping screws; one on the back and one on the
bottom of the instrument both marked with a red tag. - Remove the foam insert.
- Save the shipping screws, red tags, and foam insert with the shipping crate for future
shipping and relocation.
Note: It is important that the shipping screws are removed before plugging in the
VolumeCheck. Failure to do so may cause damage to the instrument.
- The VolumeCheck models have two shipping screws; one on the back and one on the
- Place your instrument on a vibration-free level bench top.
- Attach the power cord to the VolumeCheck and a power outlet.
- Attach a USB cable to the VolumeCheck and its computer.
- Turn on the VolumeCheck using the power switch near the power connector.