This screen is found under Settings Menu, then Ultrasound Scanning Options. This allows setting the scanning hardware parameters and error handling options.
- Rescan if the sample is misread. Recommended: enabled
- If well distance measurement is out of range the VolumeCheck will rescan the well position.
- Minimum valid distance (mm): recommended = 10mm
- Maximum valid distance (mm): recommended = 110mm
- If well distance measurement is out of range the VolumeCheck will rescan the well position.
- Rescan if Standard Deviation is higher than: recommended: enabled
- Error handling option can be set to facility requirements. recommended = 0.05
- Maximum number of retries: applies to rescans of misreads & standard deviation. Recommended = 2
- Select best readings from all scans in case retries are performed. Recommended: enabled