Advanced operations - Error Handling Options

scanner options

 

This screen is found under Settings Menu, then Ultrasound Scanning Options. This allows setting the scanning hardware parameters and error handling options.

  • Rescan if the sample is misread. Recommended: enabled
    • If well distance measurement is out of range the VolumeCheck will rescan the well position.
      • Minimum valid distance (mm): recommended = 10mm
      • Maximum valid distance (mm): recommended = 110mm
  • Rescan if Standard Deviation is higher than: recommended: enabled
    • Error handling option can be set to facility requirements. recommended = 0.05
    • Maximum number of retries: applies to rescans of misreads & standard deviation. Recommended = 2
  • Select best readings from all scans in case retries are performed. Recommended: enabled