Error recovery - 7.8 Vial fails to drop into the waste bin

Danger: Wear appropriate gloves and eye protection when handling waste vials. 

  1. Use the 'System Availability' options in the Management tab in Director's Service|Settings to take Transport offline.
  2. Look for the vial in the waste drop pipe, and in the Router arm. 
  3. If you locate the vial in the transport pipe, use the blow valve control on the Router page in Director to drop the vial into the bin. First align the Router arm under the port the bin is connected to, then operate the blow valve.
  4. Use the Vials page in Director to update the vial's location to Not in System.
  5. Re-enable transport so that the system restarts. 