Installation - Pre-installation requirements

Space

Operating dimensions  
VolumeCheck 100 29.5cm W x 68cm D* x 32cm H (11.5" x 26.5" x 12.5")
VolumeCheck 384 29.5cm W x 68cm D* x 32cm H (11.5" x 26.5" x 12.5") 

*depth includes cable clearance, and tray extension

vc dims

VolumeCheck weighs 15kg (33.25lbs).

Environment

  • 10°C - 40°C
  • 10-90% RH
  • A vibration-free bench top space with a level surface is recommended

Computer

  • Windows 10/11 PC or laptop, 32-bit / 64 bit, with monitor, keyboard and mouse
  • RAM: 2 Gb minimum
  • 1 USB Port
  • Administrator privileges needed for installation

Power

VolumeCheck requires a 110-220 VAC 60Hz electrical supply.